Dimensional error in selective laser sintering and 3D-printing of models for craniomaxillary anatomy

June 28th, 2008 | by admin |

Dimensional error in selective laser sintering and 3D-printing of models for craniomaxillary anatomy reconstruction.

BACKGROUND: Selective laser sintering (SLS) and three-dimensional printing (3DPtrade mark) are rapid prototyping (RP) techniques to fabricate prototypes from biomedical images. To be used in maxillofacial surgery, these models must accurately reproduce the craniofacial skeleton. PURPOSE: To analyze the capacity of SLS and 3DPtrade mark models to reproduce craniomaxillary anatomy and their dimensional error. MATERIAL: Dry skull, helical computed-tomography images, SLS and 3DPtrade mark prototypes, and electronic calliper. METHODS: Tomographic images of a dry skull were manipulated with the InVesalius biomedical software. Prototypes were fabricated using SLS and 3DPtrade mark techniques. Ten linear measurements were made on the models and compared with corresponding dry skull measurements (criterion standard) carried out with an electronic calliper. RESULTS: We observed a dimensional error of 2.10 and 2.67% for SLS and 3DPtrade mark models, respectively. The models satisfactorily reproduced anatomic details, except for thin bones, small foramina and acute bone projections. The SLS prototypes showed greater dimensional precision and reproduced craniomaxillary anatomy more accurately than the 3DPtrade mark models. CONCLUSION: Both SLS and 3DPtrade mark models provided acceptable precision and may be useful aids in most maxillofacial surgeries.

Silva DN, Gerhardt de Oliveira M, Meurer E, Meurer MI, Lopes da Silva JV, Santa-Bárbara A.

Department of Surgery (Head: Prof. Dr. Helena Willhelm de Oliveira), Pontifícia Universidade Católica do Rio Grande do Sul (PUCRS), Porto Alegre, RS, Brazil.

Post a Comment

page 70 page 140 page 210 page 280 page 350 page 420 page 490 page 550 page 590 page 690 page 790